Datasheet

10241-eng Rev. 09
724 Temperature Calibrator
80PK-3A
Features
The Fluke 724 Temperature Calibrator is
a powerful yet easy-to-use calibrator. Use
the measure and source functions to test
and calibrate almost any temperature
instrument.
Easy to read dual display lets you view •
input and output simultaneously
Measure RTDs, thermocouples, •
ohms, and volts to test sensors and
transmitters
Source/simulate thermocouples, RTDs, •
volts, and ohms to calibrate transmitters
Perform fast linearity tests with 25% and •
100% steps
Execute remote tests with auto step and •
auto ramp
Power transmitters during test using •
loop power supply with simultaneous
mA measurement
Store frequently-used test setups for •
later use
Fluke 724
Included Accessories
TL75 Test Leads, AC72 Test Clips, one pair of
stackable test leads.
Ordering Information
Fluke 724 Temperature Calibrator
Recommended Accessories
80PK-25TL81ATL220C25
Maximum voltage: 30V
Operating temperature: -10°C to 55°C
Safety: CSA C22.2 No. 1010.1:1992
EMC: EN50082-1:1992 and EN55022:
1994 Class B
Size (HxWxD): 200 mm x 96 mm x 47 mm
Weight: 0.65 kg
Battery: Four AA alkaline batteries
Battery life: 25 hours typical; battery door
Three Year Warranty
Function
Measure or Source
Range Resolution Accuracy Notes
Voltage 0 to 100 mV
0 to 10V (source)
0 to 30V (measure)
0.01 mV
0.001 V
0.001 V
0.02% Rdg + 2 LSD Max load, 1 mA
mA (measure) 0 to 24 mA 0.001 mA 0.02% Rdg +2 LSD Max load,
1000Ω
mV -10.00 mV to +75.00 mV 0.01 mV 0.025% of range +
1 LSD
Resistance 0Ω to 3200Ω (measure)
15Ω to 3200Ω (source)
0.01Ω to 0.1Ω 0.10Ω to 1.0Ω
Loop Supply 24V DC N/A 10%
Thermocouples J, K, T, E, L, N, U 0.1 ºC to 0.7 ºC
Thermocouples B, R, S 1 ºC to 1.4 ºC
RTDs Ni120 (672)
Pt 100, 200, 500, 1000 (385)
Pt 100 (3916)
Pt 100 (3926)
0.1 ºC to 0.2 ºC
Simultaneous Function
Capability
Channel A Channel B
24.000 mA DC M
24.000 mA DC with 24V loop
supply
M
100.00 mV DC M or S
30.000V DC Measure M
20.000V DC Measure 10.000V
DC Source
M or S
0 to 3200 Ohms M or S
Thermocouple J, K, T, E, R, S, B,
L, U, N
M or S
RTD Ni120; Pt100 (3926); Pt100
(JIS); Pt100, 200, 500, 1000 (385)
M or S
Specifications
M = Measure S = Source/Simulate
The temperature solution

Summary of content (1 pages)