® i1000s AC Current Probe for Oscilloscopes Users Manual (English)
LIMITED WARRANTY AND LIMITATION OF LIABILITY Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is one year and begins on the date of shipment. Parts, product repairs, and services are warranted for 90 days.
Table of Contents Title Page Introduction ...............................................................................1 Box Contents .............................................................................2 Safety Information .....................................................................2 Safety Specifications..............................................................4 Instrument Compatibility...........................................................4 Electrical Specifications .........
i1000s Users Manual ii
List of Tables Table Title Page 1. 2. 3. Symbols................................................................................... 4 Input Ranges and Accuracy..................................................... 6 Maximum Phase Shift ............................................................. 6 List of Figures Figure 1. 2. 3. Title Page i1000s AC Current Probe ........................................................ 2 Typical Response Curves ......................................................
i1000s Users Manual iv
Fluke i1000s AC Current Probe Introduction The Fluke i1000s is a clamp-on ac current probe designed to expand oscilloscope applications in industrial and power environments. The Current Probe (shown in Figure 1) provides the following features: • Ideal for measuring distorted current waveforms and harmonics. • Allows accurate measurement of currents from 100 mA to 1000 A rms, 5 Hz to 100 kHz without breaking into the circuit.
i1000s Users Manual Tactile Barrier aef01f.eps Figure 1. i1000s AC Current Probe Box Contents The shipping container includes the probe, this users manual, a Quick Reference Card, and a product registration card. Safety Information Read the following safety information carefully before attempting to operate or service the Current Probe. In this manual, a WWarning identifies conditions and actions that pose hazard(s) to the user.
AC Current Probe Safety Information WWarning • To avoid injury and ensure safe operation of the probe, follow all safety precautions detailed in this manual. • Never use the probe on circuits rated higher than 600 V in measurement category III (CAT III). Use extreme caution when clamping around uninsulated conductors or bus bars.
i1000s Users Manual Table 1. Symbols , May be used on HAZARDOUS LIVE conductors. T Equipment protected by double or reinforced insulation. W Risk of Danger. Important Information. See Users Manual. X Risk of Electric Shock. P Complies with EN/IEC 61010-1 2nd Edition & EN/IEC 61010-02-032 J Earth Ground B AC (Alternating Current) The i1000s complies with IEC Publication 61010-2-032. Follow all warnings to ensure safe operation.
AC Current Probe Electrical Specifications Electrical Specifications Current Range: 100 mA to 1000 A ac rms (2000 A instantaneous peak). WWarning To avoid potential thermal burns from the jaws when measuring currents equal to or greater than 800 A ac rms in the ambient temperature range of 30 to 50° C (86 to 122° F), limit the measurement cycle to a measurement time of 5 minutes or less, followed by a cooling time of 15 minutes or more. Output Signal: mV output signal (2 V peak max).
i1000s Users Manual Table 2.
AC Current Probe Mechanical Specifications Influence of Adjacent Conductor: < 1.0 mA/A ac Influence of Conductor Position in Jaw Opening: < 0.5 % of reading from 10 Hz to 5 kHz < 4.0 % of reading from 5 to 40 kHz < 10.0 % of reading from 40 to 100 kHz. Operating Temperature: -10 °C to +50 °C (14 °F to 122 °F); 100 mA to 800 A ac rms continuous, 800 to 1000 A ac rms for 5 minutes On, 15 minutes Off.
i1000s Users Manual Typical Response Curves Typical response curves are shown in Figure 2. 1 mV/A Range 1.00 Accuracy in % 0.00 -1.00 -2.00 -3.00 -4.00 @ 10 A @ 50 A -5.00 10 100 1000 10,000 100,000 10,000 100,000 Frequency (Hz) 5.00 0.00 Phase in -5.00 -10.00 -15.00 -20.00 -25.00 -30.00 @ 10 A -35.00 @ 50 A -40.00 10 100 1000 Frequency (Hz) aef07f.eps Figure 2.
AC Current Probe Typical Response Curves 10 mV/A Range Accuracy in % 5.00 0.00 -5.00 -10.00 -15.00 -20.00 @1A -25.00 @5A -30.00 10 100 1000 10,000 100,000 10,000 100,000 Frequency (Hz) 10.00 0.00 Phase in -10.00 -20.00 -30.00 -40.00 -50.00 -60.00 @1A @5A -70.00 10 100 1000 Frequency (Hz) aef08f.eps Figure 2.
i1000s Users Manual 100 mV/A Range 5.00 Accuracy in % 0.00 -5.00 -10.00 -15.00 -20.00 -25.00 -30.00 -35.00 @ 0.1 A -40.00 @ 0.5 A -45.00 10 100 1000 10,000 100,000 10,000 100,000 Frequency (Hz) 40.00 Phase in 30.00 20.00 10.00 0.00 -10.00 -20.00 -30.00 @ 0.1 A -40.00 @ 0.5 A -50.00 10 100 1000 Frequency (Hz) aef09f.eps Figure 2.
AC Current Probe Operation Operation Use the following procedure to make a measurement: 1. Connect the i1000s Current Probe to the desired input channel on the oscilloscope. 2. On the Current Probe, select the least sensitive range (1 mV/A). See Figure 1 for selector switch location. 3. On your oscilloscope, select an appropriate range. Use a 1:1 probe setting. Note If using the Fluke ScopeMeter 123, use this procedure for step 3: 4. 1. Press 2.
i1000s Users Manual 4 2 3 1000 100 10 1 CHANNEL A aef02f.eps Figure 3.
AC Current Probe Measurement Considerations Measurement Considerations Observe the following guidelines for positioning the Current Probe jaws: • • • Carefully move the probe to center the conductor inside the jaws. Make sure the probe is perpendicular to the conductor. If possible, avoid measurements close to other current-carrying conductors. On the Current Probe, the 1 mV/A range offers the best accuracy and least phase shift.
i1000s Users Manual Replacement Parts When requesting replacement parts, use the following part numbers. • Cable, 600 V rms, Safety Insulated: 935036 • Fluke i1000s Quick Reference Card: 1575200 • Fluke i1000s Calibration Manual: 1574933 Service For service, pack the Current Probe securely in its original shipping container and forward it, postage paid, to the nearest service. Include a description of the malfunction. Fluke assumes no responsibility for damage in transit.