User Manual

Battery Replacement
Observe the polarity markings inside the battery compartment.
+
+
Do not
overtighten screw
Disconnect probes
from Voltage
Switch OFF
2 AA
+
+
ex4.eps
Replacement Parts
Test Lead Assembly.
Replace only with Fluke double-insulated leads (
)
See "Accessories" for available probes.
PN 648029
Battery Door PN 1626588
Battery Door Screw PN 1618578
Accessories
H5 Belt Holster
TP1 Probe Set, Flat-Blade
TP4 Probe Set, 4 mm round
Service Information Sheet
, PN 686953
To contact Fluke, call:
USA: 1-888-99-FLUKE (1-888-993-5853)
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +31 402-678-200
Japan: +81-3-3434-0181
Singapore: +65-738-5655
Anywhere in the world: +1-425-446-5500
Visit Fluke’s Web site at www.fluke.com.
Specifications
Specifications subject to change without notice.
Calibration: One-year calibration cycle.
Maximum Voltage Between any Terminal and Earth Ground:
T5-600: 600 V rms, Overvoltage Category III; Pollution Degree 2
T5-1000: 1000 V rms, Overvoltage Category III, also 600 V rms
overvoltage Category IV, Pollution Degree 2
Maximum Voltage between Current Fork and Earth Ground: T5-600
and T5-1000: 1000 V rms, Overvoltage Category III, also 600 V rms,
Overvoltage Category IV, Pollution Degree 2, (This dual rating applies
to the current fork only.)
Temperature: Operating: -10
°
C to +50
°
C (14
°
F to 122
°
F);
Storage: -30
°
C to +60
°
C (-22
°
F to +140
°
F)
Altitude: Operating: 2000 m (6562 ft); Storage: 10,000 m (32808 ft)
Relative Humidity: 0 % to 95 %, 5
°
C to 30
°
C (41
°
F to 86
°
F); 0 % to
75 %, 30
°
C to 40
°
C (86
°
F to 104
°
F); 0 % to 45 %, 40
°
C to 50
°
C
(104
°
F to 122
°
F)
Battery Type and Life: AA (2); 360 hours continuous with alkaline;
125 hours continuous with zinc chloride
Shock, Vibration: 1 m drop at -10
°
C to 50
°
C (14
°
F to 122
°
F) per
ANSI/ISA-S82.01-1994 and EN 61010-1: 1995. Random vibration per
MIL-PRF-28800F for a Class 2 instrument (5 Hz to 55 Hz, 3 g
maximum).
Surge Protection: T5-600 6 kV per IEC 61010;
T5-1000 8 kV per IEC 61010
Enclosure Rating: IP 52 per IEC 60529, no vacuum applied
RF Field Specification: 0.5 % full scale + (specified accuracy) at
3 V / m
Safety: Complies with ANSI/ISA-S82.01-94, UL classified to IEC
61010, CSA/CAN C22.2 No.1010.1-92, and EN 61010-1: 1995.
EMC: EN 61326
Certifications:
In accordance with
IEC 61010-1 54CJ