060AM-NM-LRVAR Local, Remote, and Flasher Alarms Operators Manual March 2005 Manual No. 1060AMNMLRVAR-1 Rev. 1 ©2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A.
Fluke Biomedical Radiation Management Services 6045 Cochran Road Cleveland, Ohio 44139 440.498.2564 www.flukebiomedical.
Table of Contents Section 1: 1.1 Introduction................................................................................................ 1-1 Procedures, Warnings and Cautions ........................................................... 1-1 Section 2: 2.1 2.2 2.3 2.4 2.5 2.6 2.7 Instructions ................................................................................................ 2-1 General Description .....................................................................................
(Blank Page)
Introduction Procedures, Warnings and Cautions 1 Section 1 Introduction 1.1 Procedures, Warnings and Cautions • • • • • • • • • • • The equipment described in this manual is intended to be used for the detection and measurement of ionizing radiation. It should be used only by persons who have been trained in the proper interpretation of its readings and the appropriate safety procedures to be followed in the presence of radiation.
1060AM-NM-LRVAR Operators Manual (Blank Page)
Instructions General Description 2 Section 2 Instructions 2.1 General Description Refer to 1060-005 drawing located in Appendix A . The Fluke Biomedical local , remote , and flasher alarms are used in conjunction with the Victoreen Model 1060AM-NM-LR. The operating range of this system is 0.01 mR/hr to 1R/hr. The red halogen lamps and the flashing yellow sign provides a remote (75ft) and a local visual alarm indication.
1060AM-NM-LRVAR Operators Manual Electrical Requirements 120 VAC 50/60 Hz 120 VAC 50/60 Hz 120 VAC 50/60hz 2.3 Receiving Inspection Upon receipt of the unit: 1. Inspect the carton(s) and contents for damage. If damage is evident, file a claim with the carrier and notify Fluke Biomedical at 440.248.9300. 2. Remove the contents from the packing material. 3. Verify that all items listed on the packing list have been received and are in good condition.
Instructions Installation 2 2.5 Installation WARNING To prevent electrical shock, ensure that power is disconnected before installing the alarms Installation of the 1060am-nm-lrvar consists of mounting the unit , providing an electrical interface , and performing an operational check. The 1060-024,1060-034, and 1060-026 have four mounting holes inside each chassis with a 0.19 diameter clearance. 1 Remove four sheet metal screws from the case assembly . (1060-024 and 1060-034).
60AM-NM-LRVAR Operators Manual 2.6 Electrical interface The Local alarm (1060-024) requires power connections as follows: (See appendix A for the 1060-13 and 1060-024 drawings) Terminate the input power to TB2-1(L1) of the Local alarm assembly at the end of this installation . 1. Earth ground to chassis lug items 17 and 18 in view A-A of the local alarm assembly. 2. The L2 neutral connection (white) to TB1-8. 3. Connect the 75 foot power interface (male end) cables (1060-022 ) to both J1 and J2 .
Instructions Operation 2 2.7 Operation The red beacons and red leds are ON . The flasher alarm is blinking . Connect the free end of the I/O (1060ds-24) cable to the model 1060AM-NM-LR I/O connector. Connect the 1060AM-NM-110 (12 volt 500 mA) power cube connector to the 1060AMNM-LR power input . The red beacons turn off and the green ones are on .The green LED’s are also on. The flasher alarm is now off .
1060AM-NM-LRVAR Operators Manual (Blank Page)
Maintenance Maintenance and Cleaning 3 Section 3 Maintenance and Cleaning WARNING To prevent injury, do not remove or insert lamp when unit is energized . CAUTION To prevent damage to the lens, do not use abrasive materials or cleaners Refer the parts lists for the required type of lamp After disconnecting power, replace the lamp as follows: 1. Insert a small flat blade screwdriver between the INSIDE locking mechanism and the lens.
1060AM-NM-LRVAR Operators Manual 1. Slide the acrylic yellow panel out from either side of the chassis . CAUTION To prevent property damage and injury, do not touch glass with bare fingers. Grasp glass with a soft clean cloth or with packaging supplied with the replacement lamp. 2 . Gently push in on the lamp, twist and pull up to remove the lamp from the socket. Install the new lamp(s) by aligning the connector on the base of the tube with its mating socket, then carefully press down into the socket.
Appendix Applicable Drawings and Parts List A Appendix A Applicable Drawings and Parts List 4.
Fluke Biomedical Radiation Management Services 6045 Cochran Road Cleveland, Ohio 44139 440.498.2564 www.flukebiomedical.