FUNCTION REFERENCE 5 - 9
If switching from one mode to another results in an unsatisfactory display, press 
the yellow key a second time to return to the original situation.
Key sequence:
Toggle key to switch between analog mode and digital 
mode.
AUTO RANGE
Description:
The AUTO RANGE function automatically selects the input sensitivity of the 
vertical channels and time/div of the main timebase to obtain an optimum display 
of the input signal(s). AUTO RANGE is a continuous function: vertical channel or 
time/div settings are adapted if the input signal changes.
The AUTO RANGE function of the vertical channels automatically selects the 
input sensitivity. The result is that the input signal is displayed with 2 to 6.4 
divisions of amplitude. The amplitude is between 1 and 3.2 divisions with 
DISPLAY WINDOWS on. AUTO RANGE can be turned on separately for each 
vertical channel. AUTO RANGE functions at input sensitivities between 5V and 
50 mV/div. Input coupling is put to ac if AUTO RANGE is switched on: dc or 
ground coupling can be selected manually. AUTO RANGE is switched off after 
operation of the AMPL key, the AUTOSET key, the channel’s AUTORANGE key, 
channel ADD key or changing Acquisition length.
The AUTO RANGE function of the main timebase (MAIN TB) adjusts the 
timebase automatically so that 2 to 6 waveform periods are displayed. To turn off 
AUTO RANGE, press the TIME/DIV key, the AUTOSET key, or the main timebase 
AUTORANGE key.
In VARiable timebase mode the AUTO RANGE function tries to keep the 
displayed trace unchanged.
Key sequence:
Toggle key to switch the AUTO RANGE function of main 
timebase and vertical channels.
ANALOG
AUTO RANGE