Musical Instrument User Manual
FUNCTION REFERENCE 5 - 11
AUTOSET SEQUENCE
Description:
If front panel settings are stored in memory locations ‘s1 to s5’ and ‘s6’ is cleared, 
then the range of ‘s1 to s5’ becomes a sequence of front panel settings. Such a 
sequence can be used as (part of) a step-by-step testing procedure. The two 
following methods can be used to quickly step through such a sequence. This 
saves time by not without having to recall each setting from a menu:
- Programming the AUTOSET key to be used as the ‘recall next setup’ key. This 
selection is done in the UTILITY >> AUTOSET menu.
- Using a special probe with a COMMAND switch. Two probes with COMMAND 
switch are delivered with each oscilloscope as standard accessory. The 
appropriate mode for this probe command can be selected in the UTILITY >> 
PROBE menu.
To return to the ‘normal’ oscilloscope mode, press the STATUS and TEXT OFF 
key simultaneously to recall the STANDARD SETUP.