Musical Instrument User Manual
4 - 46 HOW TO USE THE INSTRUMENT
CURSORS VOLT
- Use the softkeys, to select the amplitude cursors (’=’ intensified).
Two amplitude cursors will appear in the display.
The menu permits the channel to be selected for which the amplitude cursors 
apply.
When channel 1 is the only channel displayed, ’ch1’ is the only selection.
-
Turn the TRACK 
control to move both 
cursors, and turn the 
Delta (
∆
) control to 
move only the 
∆
cursor
.
CURSORS AMPLITUDE & TIME
Amplitude and Time cursors can be displayed at the same time. When both 
amplitude and time cursors are displayed, you must select which cursors are 
affected by the TRACK and ∆ controls. This selection is made by toggling the 
softkey ’CONTROL’.
Channel selection applies only to the voltage cursors. The time cursor always 
applies to all channels.
To use both types of cursors at the same time, do the following:
- Press the softkey next to ’= || #’ so that ’#’ is intensified.
- Press the CONTROL softkey so that ’=’ is intensified.
- Turn the TRACK control and observe that only the amplitude cursors 
move.
- Press the CONTROL softkey so that ’||’ is intensified.
- Turn the TRACK control and observe that only the time cursors move.
CURSORS
READOUT
ST6731
9303
CH1
200mV
MTB 200us
ch1
on off
#
-
ch1
-
TRACK
∆
ch1: ∆T=
s ∆V=360mV