Musical Instrument User Manual
HOW TO USE THE INSTRUMENT 4 - 49
4.11 MEASUREMENT FUNCTIONS
When operating in the 
analog
 mode, you can use the cursors to measure 
amplitude and time data.
When operating in the 
digital
 mode, the scope has an extensive set of 
fully 
automated
 amplitude and time measurement functions. You can select two 
measurements to be performed simultaneously. These measurements are 
updated automatically, so that when signals change, the measured values 
automatically changed at the same time. The read- outs are displayed in the 
upper section of the screen.
STANDARD SETTING
Start with the standard setting.This ensures a correct start.
- Simultaneously press the STATUS and TEXT OFF keys.
- Connect the Probe Adjust signal to channel 1.
The Probe Adjust signal now supplied to the input is a square wave with a lower 
level of 0V and a top level of 600 mV.
-Press AUTOSET.