User's Manual
PN 686953
March 1998 Rev.1, 9/01
1998-2001 Fluke Corporation. All rights reserved. Printed in U.S.A. Product names are trademarks of their respective companies.
T5-600/T5-1000
Electrical Tester
Service Information
Introduction
This service information sheet provides the following information for the T5-600 and T5-1000 Electrical
Testers (hereafter referred to as "the tester").
• Safety information
• Parts and service information
• Specifications
• Cleaning procedure
• Required equipment
• Performance tests
• Parts and accessories list
• Battery replacement procedure
For operating instructions, refer to the T5-600/T5-1000 Electrical Tester Instruction Sheet.
Safety Information
Warning
To avoid possible electric shock or personal injury, follow these guidelines:
• Do not use the tester if it is damaged or operating abnormally. Protection
may be impaired.
• Before each use:
• Make sure the battery door is closed and latched.
• Inspect the tester and test leads. Look for cracks, missing plastic, exposed
metal, or damaged insulation. Replace damaged test leads before using the
tester.
• Verify the tester’s operation by measuring a known voltage.
• Replace the batteries as soon as the low battery indicator (M) appears.
• Do not use the tester around explosive gas, vapor or dust.
• Do not apply more than the rated voltage, as marked on the tester, between
terminals or between any terminal and earth ground.
• Refer servicing to qualified personnel.
• Use caution when working above 30 V ac rms, 42 V ac peak, or 60 V dc.
• When using the probes, keep your fingers behind the finger guards on the
probes.
• Connect the common test lead before you connect the live test lead.
Disconnect the live test lead first.
• The display will not display hazardous voltages in display hold. The voltage
indicator LED (
) continues to operate.
®