User's Manual

PN 686953
March 1998 Rev.1, 9/01
1998-2001 Fluke Corporation. All rights reserved. Printed in U.S.A. Product names are trademarks of their respective companies.
T5-600/T5-1000
Electrical Tester
Service Information
Introduction
This service information sheet provides the following information for the T5-600 and T5-1000 Electrical
Testers (hereafter referred to as "the tester").
Safety information
Parts and service information
Specifications
Cleaning procedure
Required equipment
Performance tests
Parts and accessories list
Battery replacement procedure
For operating instructions, refer to the T5-600/T5-1000 Electrical Tester Instruction Sheet.
Safety Information
Warning
To avoid possible electric shock or personal injury, follow these guidelines:
Do not use the tester if it is damaged or operating abnormally. Protection
may be impaired.
Before each use:
Make sure the battery door is closed and latched.
Inspect the tester and test leads. Look for cracks, missing plastic, exposed
metal, or damaged insulation. Replace damaged test leads before using the
tester.
Verify the tester’s operation by measuring a known voltage.
Replace the batteries as soon as the low battery indicator (M) appears.
Do not use the tester around explosive gas, vapor or dust.
Do not apply more than the rated voltage, as marked on the tester, between
terminals or between any terminal and earth ground.
Refer servicing to qualified personnel.
Use caution when working above 30 V ac rms, 42 V ac peak, or 60 V dc.
When using the probes, keep your fingers behind the finger guards on the
probes.
Connect the common test lead before you connect the live test lead.
Disconnect the live test lead first.
The display will not display hazardous voltages in display hold. The voltage
indicator LED (
) continues to operate.
®