user manual
Table Of Contents
- 45 Dual Display Multimeter
- 1. Introduction
- 2. Getting Started
- 3. Operating the Meter From the Front Panel
- 4. Applications
- 5. Operating the Meter Using the Computer Interface
- Introduction
- Preparing the Meter for Operations via the RS-232 Interface
- Preparing the Meter to be Operated via IEEE-488 Interface
- Getting Started With An Installation Test
- How the Meter Processes Input
- How the Meter Processes Output
- Triggering Output
- Service Requests (IEEE-488 Only and Status Registers
- Computer Interface Command Set
- IEEE-488 Capabilities and Common Commands
- Function Commands and Queries
- Function Modifier Commands and Queries
- Range and Measurement Rate Commands and Queries
- Measurement Queries
- Compare Commands and Queries
- Trigger Configuration Commands
- Miscellaneous Commands and Queries
- RS-232 Remote/Local Configurations
- Sample Program Using the RS-232 Computer Interface
- Sample Programs Using the IEEE-488 Computer Interface
- 6. Maintenance
- Appendices
45
Users Manual
6-6
Table 6-2. Performance Tests for Volts, Diode Test, Ohms, and Frequency, Functions (cont.)
Display
Function Range (Rate)
Input Level
Frequency
Min Max
Using decades of 3:
300 Ω
3 kΩ
30 kΩ
300 kΩ
3 MΩ
30 MΩ*
300 MΩ*
Short
300 Ω
short
3 kΩ
30 kΩ
300 kΩ
3 MΩ
30 MΩ
300 MΩ
Using decades of 1.9:
300 Ω
3 kΩ
30 kΩ
300 kΩ
3 MΩ
30 MΩ*
300 MΩ*
Short
190 Ω
short
1.9 kΩ
19 kΩ
190 kΩ
1.9 MΩ
19 MΩ
190 MΩ
Using decades of 1:
300 Ω
3 kΩ
3 kΩ
30 kΩ
300 kΩ
3 MΩ
30 MΩ*
300 MΩ*
Short
100 Ω
short
1 kΩ
10 kΩ
100 kΩ
1 MΩ
10 MΩ
100 MΩ
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
—
0.00
299.83
0.0000
2.9983
29.983
299.83
2.9980
29.922
294.0
0.00
189 88
0.0000
1.8988
18.988
189.88
1.8987
18.949
186.2
0.00
99.93
0.0000
0.9993
9.993
99.93
0.9992
9.972
98.0
0.04
300.19
0.0002
3.0017
30.017
300.17
3.0020
30.078
306.0
0.04
190.14
0.0002
1.9012
19.012
190.12
1.9013
19.051
193.8
0.04
100.09
0.0002
1.0007
10.007
100.07
1.0008
10.028
102.0
FREQ 10/100 kHz 0.1 V to 10 V 10 kHz 9.9949 10.006
* Optional test points that can be used if standards are available.