Supplement Title: VR1710 User Print Date: October 2007 Revision/Date: Supplement Issue: 5 Issue Date: 7/13 Page Count: 7 This supplement contains information necessary to ensure the accuracy of the above manual. This manual is distributed as an electronic manual on the following CD-ROM: CD Title: CD Rev. & Date: CD PN: © 2007-2013 Fluke Corporation. All rights reserved. Printed in the U.S.A.
VR1710 Users Manual Supplement Change #1, 43993 On page 9, under Electrical Specifications, Change: Harmonic amplitude uncertainty ±0.1 % relative to fundamental RMS value To: Harmonic amplitude uncertainty 10/07 ±0.
Manual Supplement VR1710 Users Change #2 On page 5, under Using the Recorder, Change: Loading the USB Driver To: USB Driver Software Installation Delete: Delete: To load the VR1710 USB driver: 4. Restart the PC to conclude driver installation. On page 6, under Connecting the Recorder for the First Time, after the second step 2, add: 3. When the New Hardware Wizard appears again, follow the procedure above to complete USB driver software installation.
VR1710 Users Manual Supplement Change #3 On page 7, replace step 3 and Figure 3 with: 3. Click on (VR1710 Settings). From the Settings screen, you can specify the following: 1/10 Recording interval from 1 second to 20 minutes. To change the interval setting, recorded data must first be erased. Dip and Swell thresholds from 0 V to 300 V. Circular memory on or off. Circular memory on will continuously record, writing over old data when memory is full.
Manual Supplement VR1710 Users Transient sensitivity from 2 V to 20 V. To detect transients, a typical waveform is first determined through sampling. This waveform changes over time as the input waveform changes. An envelope around the typical waveform is created based on the sensitivity setting. If any point of a sample waveform falls outside of the envelope, the waveform is recorded as a transient. Note VR1710 firmware version 1.15 or higher is required for circular memory and transient settings.
VR1710 Users Manual Supplement fbx003.bmp Figure 3.
Manual Supplement VR1710 Users Change #4, 57565 On page 9 under Electrical Specifications, Change: RMS voltage measurement range 70 V to 300 V RMS voltage measurement range 70 V to 270 V continuous, 270 V to 300 V <50 ms transient.
VR1710 Users Manual Supplement Change #5, 66208 On page 3, add the following to the Symbols Table: Conforms to relevant South Korean EMC Standards. On page 10, remove the IP Rating, Vibration requirements, Drop test requirements, Measurement category rating, Safety, and EMC from the Environmental and Safety Specifications. Replace with: Safety ............................................... IEC 61010-1: Overvoltage Category II, Measurement 300 V CAT II, Pollution Degree 2 IP Rating ..................