4 1/10
Transient sensitivity from 2 V to 20 V. To detect
transients, a typical waveform is first determined through
sampling. This waveform changes over time as the input
waveform changes. An envelope around the typical
waveform is created based on the sensitivity setting. If
any point of a sample waveform falls outside of the
envelope, the waveform is recorded as a transient.
Note
VR1710 firmware version 1.15 or higher is required for
circular memory and transient settings.