User's Manual
Manual Supplement VR1710 Users
6 2/11
Change #4, 57565
On page 9 under Electrical Specifications,
Change:
RMS voltage measurement
range
70 V to 300 V
To:
RMS voltage measurement
range
70 V to 270 V continuous, 270 V
to 300 V <50 ms transient.
On page 10 under General Specifications
,
Change:
Power requirements 70 V to 300 V
To:
Power requirements 70 V to 270 V