5.5 Start-up and operation
a Set points
b Actual values
c Adjustment buttons: left (-), right (+)
d Function buttons
For setting the sieving time, interval time, and amplitude, there is a pair
of buttons for each setting. The left button decreases the set point and
the right button increases it. Hold one of the buttons down to scroll
through the numbers more quickly.
5.5.1
Switching on
Switch on the machine with the main switch (which is located on the
underside of the control device). Zero will appear in each display.
5.5.2 Time
The sieving time is set in minutes in the set point / actual time display.
The maximum value is 99 minutes. After pressing "Start", the display
switches to the actual value display, which counts down the remaining
sieving time. When the set point has elapsed, the sieving process ends. If
no sieving time (set point 00) is provided, the machine will run without a
time limit.
Using the device