Owner's manual

6.9.1 Changing the time unit in setup mode
1. When the device has been switched off, press and hold the STOP
button on the front control panel.
2. Switch on the device with the main switch (8) on the back of the
device.
3. If POWER SUPPLY is flashing, the device is in setup mode. If
POWER SUPPLY is not flashing, repeat the procedure.
4. To perform changes, press the right "+" button (y) in the TIMER
field:
Time unit, hours and minutes ®
display: -
Time unit, minutes and seconds ® display: 1 (factory setting)
5. To save and end setup mode, press the STOP button.
6.10
Reverse mode
®Press the REVERSE button. If the REVERSE button is illuminated,
reverse mode is activated.
After the selected running time expires, the mill will change its direction
of rotation. For this, REPETITIONS has to be at least 1.
"Reverse" is selected when mixing dry samples or suspensions, for
example.
Using the device