Manual

8 Cleaning
8.1
Grinding elements
Clean the grinding bowl and grinding balls each time after using them:
e.g. clean them under running water using a brush and a commercially
available cleaning agent.
Add sand and water until the grinding bowl is about half full and "grind
to clean" for 2 to 3 minutes in the laboratory planetary mill.
Cleaning with an ultrasonic cleaner is permitted.
For sterilisation in the heat cabinet, only heat the grinding elements up
to 120°C.
Grinding elements made of agate may only be heated up to 100°C.
CAUTION!
Cool grinding elements made of agate, sintered corundum,
zirconium oxide and silicon nitride slowly and carefully.
Do not heat the grinding elements in a microwave under
any circumstances (heating is too fast).
They must never be exposed to thermal shocks as this
could cause irreparable damage to the parts They will fall
apart in an explosive manner.
8.2 Mill
DANGER!
Mains voltage!
Before beginning with cleaning work, disconnect the
mains plug and protect the device against being unin-
tentionally switched back on!
Do not allow any liquids to flow into the device.
Indicate cleaning work with warning signs.
Put safety equipment back into operation after
cleaning work.
The micro mill can be wiped down with a damp cloth when it is switched
off.
The grinding chamber can be cleaned thoroughly with a damp cloth by
removing the grinding chamber cover.
Cleaning
- 101 -