Manual
6.3.3.2 Transmission frequency setting options
Transmission every Switch 3 Switch 4
1 second OFF OFF
½ second ON OFF
¼ second OFF ON
automatic ON ON
1-second transmission is the default value.
For automatic transmission, the 1 / ½ / ¼ second transmission frequency
is switched over in the event of fast changes in pressure or temperature.
The data transmission frequency naturally also has an impact on the bat-
tery's service life.
1-second operation results in the longest time span.
The transmission frequency can also be checked visually: as soon as the
transmission unit is switched on, the switch flashes each time data is
transmitted.
6.3.4 Installation of the transmission unit on the EASY GTM bowl
NOTICE!
The threaded holes inside the bowl, the passage holes and
the Allen screws in the locking ring have to be cleaned thor-
oughly after each grinding to allow a safe fastening of the
EASY GTM system.
The EASY GTM system must be assembled as follows before inserting the
EASY GTM system in the micro mill:
Accessories
- 75 -