Datasheet
PP 8 FW 7333SM
Applications
■
Blood analyser
■
Patient monitors
■
Measuring equipment
■
Laboratory equipment
■
Inhalers
■
Patient lifts
Characteristics
■
Universal input 100 to 240 V AC
■
Constant voltage, current limited
■
Green LED indicator
■
Low standby power
≤ 0.3 Watts (PP 8) resp.
≤ 0.5 Watts (MPP 15 / 30)
■
Continuously short circuit proof
■
Exchangeable primary adapters (MPP 15 / 30)
Technical data
Input voltage 100 to 240 V AC (± 10%)
Input current 200 mA (PP 8), 400 mA (MPP 15),
700 mA (MPP 30)
Frequency 50 to 60 Hz
Effi ciency 80 % typ. at full load, PP 8 75%
EMC Conforms to
EN 55011, EN 55022 / B,
FCC 47 part 15, EN 61000-3-2,
EN 61000-4-2, EN 61000-4-3,
EN 61000-4-4, EN 61000-4-5,
EN 61000-4-6, EN 61000-4-11
Output voltage tolerance ± 5%
Environmental specifi cation
Operating temperature 0 to 40° C at maximum load
Storage temperature -40 to 70° C, PP 8 (-20 to 70° C)
Humidity 5 % to 95 % non condensing
Input transient susceptibility Complies with IEC 61000 requirements
Safety specifi cation
Standards Fulfi ls Class II SELV for the following
applications:
IEC 60601-1, UL 2601, VDE, CE label,
fulfi ls medical application class B / BF / CF
Reliability specifi cation
MTBF calculation 200,000 hours at maximum load
and an ambient temperature of 25° C
(in accordance with MIL-HDBK-217)
Mechanical specifi cation
Weight approx. 110 g (PP 8), 160 g (MPP 15), 255 g (MPP 30)
Plug connector AC input:
FRIWO exchangeable mains plug system:
EURO, UK, USA / Japan, Australia, IEC
PP 8: EURO, UK, USA / Japan
DC output:
Universal output plug system (page 30)
For primary adapters see page 30
MPP 15 FW 7555M
MPP 30 FW 7362M
Switchmode Power Supplies
Medical Series
with exchangeable primary adapters
(MPP 15 / 30)
All products conform to IEC 60601-1
16