Wrist Blood Pressure Monitor (Electronic Sphygmomanometer) User Manual Thank your for purchasing our product. Please read this manual carefully before using the product. Please keep the manual appropriately for future reference. This User Manual is suitable to the following models: FT-B11W, FT-B12W, FT-B13W, FT-B14W, FT-B21Y, FT-B22Y, FT-B11W-V, FT-B12W-V, FTB13W-V, FT-B14W-V, FT-B21Y-V, FT-B22Y-V, FT-B13W-UR and FT-B13W-BT.
I. General Information Intended Use and Indications For Use Fudakang Wrist Blood Pressure Monitor are non-invasive blood measurement system intended to measure the diastolic, systolic blood pressures and pulse rate of an adult individual in hospitals, hospital-type facilities and home environments. The BT series Blood Pressure Monitor with the wireless communication function that is connect to the PC or a mobile phone for record archiving and printing purpose.
Transport and storage Environment Temperature: -20~55°C Humidity: <95%RH Pressure: 86~106 kPa Electric Shock Protection Internal power unit Memory Capacity 90 sets of data including date and time See VI. Model List Inflation Automatic Inflation by internal pump Deflation Automatic speed deflation system controlled by internal electromagnetic valve.
Maintenance: This product is designed for use over an extended period of time; however, it is generally recommended that it be inspected every five years to ensure proper function and performance. The device doesn’t need to be calibrated in five years of reliable service. Modification of this equipment is allowed except change the batteries.
9. Measurement may not be possible for anyone with insufficient blood flow to the area where measurements will be taken or who suffers from a frequent irregular heartbeat. Consult your physician for advice on whether to use the device. 10. DO NOT wrap the cuff around an injured wrist. 11. DO NOT wrap the wrist cuff around a wrist in which a drip (intravenous infusion) is inserted or which is being used for blood transfusion as part of medical treatment.
5. NEVER clean the blood pressure monitor with thinners or benzene, as they may damage it. 6. DO NOT hard rub when clean the cuff. 7. DO keep the device out of reach of children, pets and insects. III.
Figure 2 – Display of LCD Note for LCD display: (1) Date: Month – Day (2) Time: Hour – Minute (3) Systolic Blood Pressure (unit: mmHg) (4) Diastolic Blood Pressure (unit: mmHg) (5) Pulse Rate (unit: beat/minute) (6) WHO Blood Pressure Classification Indicator (7) Inflation / Deflation Indicator (8) Blood Pressure Measurement Unit (9) Battery Symbol (10) Irregular heartbeat Indicator (11) Memory Record Number * Back light display function is optional. IV.
Place batteries with positive “+” and negative “-“ terminals into compartment and make sure they match the indicated terminals in the compartment. Close the battery cover by gently sliding it into the compartment and pressing it into place.. See Figure 3. Note: : When the LCD display shows “Low Battery” signal ,the batteries must be replaced for accurate readings. See Figure 4 Do not use rechargeable batteries (voltage 1.2V).
DAY signal will begin to blink. Press the “M” button to change the DAY. Each time when you press the “M” button, it will change one DAY forward. Don't keep on clicking on the 'SET' button without being released during programming. REPEAT THIS PROCESS FOR SETTING THE TIME. Use the “SET” button to change (Hours/Minutes) and the “M” button will change the numbers forward . See Figure 5. After you change the batteries, you have to readjust the date and time. Time is maintained using a 24 HOUR clock.
fitting the cuff. ★ Unless your physician recommends otherwise, use the left wrist to measure pressure. ★ The cuff should be snug but not too tight. You should be able to insert two fingers between the cuff and your wrist. See Figure 6. ★ Only use the manufacturer cuff with the main unit to ensure accurate measurement.
Figure 7 After you are in a comfortable position, press the “ON/OFF” button. The device will perform a self verification/check. During this verification/check the LCD will display all “8’s”. At the conclusion of the verification/check the LCD will display “00”. See Figure 8. If the device has voice function, it will speak out the displayed blood pressure, heart rate. If an irregular heartbeat is detected, the IRREGULAR HEARTBEAT symbol will appear and blink in the display screen. See Figure9.
★ To review other results that are in memory – Press the “M” button to scroll through previous measurements. Each time you press and release the “M” button the next oldest result will be displayed.
From the above figure, we can see the classification of blood pressure for adults is as below. The WHO BP Classification Indicating Bar would show out the blood pressure level by the color indicator.
Shut Down After measurement, press button “ON/OFF” to turn off the device. The device will be automatically power off after 1 minute of none use. Voice Function The device with voice function can speak out in the following state: The device will speak out the prompt that “keep silence to take a cuff at the same height with your heart ” when the measure begin.
● The UART port connector & Cable specification: 1. Any type of USB (micro USB, mini USB, standard USB) or serial connector is defined by the customer. 2. Cable : OD 3.5 +/- 0.1 mm , # 28 x 4 Color wires 3. Contact resistance >2 ohm 4. Insulation resistance: DC 300V 20 Mohm /10 ms The Blue Tooth function model is FT-B13W-BT. Operation Method: ★ Install the APK accordance with the communication protocol into the blue tooth signal receiving device such as mobile phone.
After mobile phone Bluetooth module searched to find the Bluetooth blood pressure monitor, it display “ClinkBlood” . Ⅱ.Press the “ClinkBlood” till “Connected” display on mobile phone screen which means successfully connected with blood pressure monitor. Ⅲ.Slide mobile phone screen to "Slave - > Host" and switch its NOTIFIED VALUES status from "Listen for notifications" to "Stop listening", which means it will be status of "Listening" once open. As shown in Figure 2 and Figure 3 Figure 2 Figure 3 Ⅳ.
The above two models have all function as normal NIBP only have different “result output “ method. V. Troubleshooting Abnormality Reason Checkout LCD shows Low Battery icon Batteries are low. Change new batteries. Shows abnormal result Cuff is not tightened properly or its position is incorrect. The arm is moved during measuring. Tighten cuff correctly and refer to “Wrist Cuff Connecting”. Stay calm, arm remains steady. Do not move during measuring.
Irregular heartbeat You can test again for light irregular heartbeat patients. It is inappropriate for serious irregular heartbeat patients to use this device. Speaking, frightened nervous or excited measurement Do not speak, take deep breath 2~3 times to relax yourself. Wrong position Adjust position; refer to “Wrist Cuff Swathing”. Some interference in inflation or wrong operation during measuring Refer to the inflation step in “Measuring process”.
STATEMENTS AND DECLARATIONS: 1. MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in the ACCOMPANYING DOCUMENTS 2. Wireless communications equipment such as wireless home network devices, mobile phones, cordless telephones and their base stations, walkie-talkies can affect this equipment and should be kept at least a distance d = 3,3 m away from the equipment. (Note.
RF emissions CISPR 11 RF emission CISPR 11 Harmonic emissions IEC 61000-3-2 Voltage fluctuations/ emissions IEC 61000-3-3 The [EQUIPMENT or SYSTEM] use RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
Power frequency (50Hz) magnetic field IEC 61000-4-8 NOTE <5% UT (>95% dip in UT) for 5 sec 3A/m 3A/m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment. UT is the a.c. mains voltage prior to application of the test level. Guidance and manufacture’s declaration – electromagnetic immunity The [EQUIPMENT or SYSTEM] is intended for use in the electromagnetic environment specified below.
b observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the [EQUIPMENT or SYSTEM]. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. Recommended separation distances between portable and mobile RF communications equipment and the [EQUIPMENT or SYSTEM].
Explanation of Symbols: LOT Symbol for batch code Symbol for manufacturer Symbol for ‘CE” Symbol for “electrical and electronic equipment” Symbol for “TYPE BF APPLIED PART” Symbol for “Follow operating instructions” IP22 Symbol for “the IP classification” Symbol for “ RF transmitters” Manufacturer: FUDAKANG INDUSTRIAL CO.,LTD Address :No.8 Yinghe Road, Yuanjiangyuan Management Zone, Changping Town, Dongguan, Guangdong China. Tel: 86-769-81098181 Software Version 1.3 Manual Version: V2.