User's Manual

RF emissions
CISPR 11
Group 1
The [EQUIPMENT or SYSTEM] use RF energy
only for its internal function. Therefore, its
RF emissions are very low and are not
likely to cause any interference in nearby
electronic equipment.
RF emission
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/ flicker
emissions
IEC 61000-3-3
Not applicable
Guidance and manufacture’s declaration
electromagnetic immunity
The [EQUIPMENT or SYSTEM] is intended for use in the electromagnetic environment specified below. The customer
or the user of
[EQUIPMENT or SYSTEM]
should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile. If
floor are covered with
synthetic material, the
relative humidity should be
at least 30%. If ESD
interfere with the operation
of equipment ,counter
measurements such as
wrist strap, grounding shall
be considered.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for
input/output
lines
Not applicable
Mains power quality should
be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV differential
mode. ±2 kV common
mode
Not applicable Mains power quality should
be that of a typical
commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
Not applicable Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
the TL-100Drequires
continued operation during
power mains interruptions, it
is recommended that the
TL-100Dbe powered from an
uninterruptible power supply
or a battery.