Arm Blood Pressure Monitor (Electronic Sphygmomanometer) User Manual Thank your for purchasing our product. Please read this manual carefully before using the product. Please keep the manual appropriately for future reference. This User Manual is suitable to the following models: FT-C21Y, FT-C22Y, FT-C23Y, FT-C24Y, FT-C11B, FT-C12B, FT-C21Y-V, FT-C22Y-V, FTC23Y-V, FT-C24Y-V, FT-C11B-V, FT-C12B-V, FT-C11B-UR and FT-C11B-BT.
I. General Information Intended Use and Indications For Use Fudakang Arm Blood Pressure Monitor are non-invasive blood measurement system intended to measure the diastolic, systolic blood pressures and pulse rate of an adult individual in hospitals, hospital-type facilities and home environments. The BT series Blood Pressure Monitor is with the wireless communication function that is connected to the PC or a mobile phone for record archiving and printing purpose.
Transport and storage Environment Temperature: -20~55°C Humidity: <95% RH Pressure: 86~106 kPa Electric Shock Protection Internal power unit Memory Capacity 2 Memory sets ; each 90 reading of data including date and time Inflation Automatic Inflation by internal pump Deflation Display Color Backlight display on LCD (Optional) Switch Life Time Contents Automatic speed deflation system controlled by internal electromagnetic valve.
3. To clean the cuff, wipe it with a moist cloth. Avoid hard rubbing as this will cause air leakages. Take care also not to get water into the air hose. Maintenance: This product is designed for use over an extended period of time; however, it is generally recommended that it be inspected every five years to ensure proper function and performance. The device doesn’t need to be calibrated in five years of reliable service. Modification of this equipment is allowed except change the batteries.
8. If you suffer from a severe problem with blood circulation in your arms, consult your physician before using the device. Failure to do so could be hazardous to your health. 9. Measurement may not be possible for anyone with insufficient blood flow to the area where measurements will be taken or who suffers from a frequent irregular heartbeat. Consult your physician for advice on whether to use the device. 10. DO NOT wrap the cuff around an elbow. DO NOT wrap over a wound. 11.
4. DO NOT attempt to disassemble the device. User can open battery cover for new battery installation. DO NOT bend the cuff or air hose excessively. 5. NEVER clean the blood pressure monitor with thinners or benzene, as they may damage it. 6. DO NOT hard rub when clean the cuff. Take care not to get water into the air hose. 7. DO keep the device out of reach of children, pets and insects. III.
Figure 2 – Display of LCD Note for LCD display: (1) Date: Month - Day (2) Time: Hour – Minute (3) Systolic Blood Pressure (unit: mmHg) (4) Diastolic Blood Pressure (unit: mmHg) (5) Pulse Rate(unit: beat/minute) (6) WHO Blood Pressure Classification Indicator (7) Inflation / Deflation Indicator (8) Blood Pressure Measurement Unit (9) Battery Symbol (10) Irregular heartbeat Indicator (11) Memory Record Number * Back light display function is optional. IV.
Close the battery cover by gently sliding it into the compartment and pressing it into place.. See Figure 3. Note: : When the LCD display shows “Low Battery” signal ,the batteries must be replaced for accurate readings. See Figure 4 Do not use rechargeable batteries (voltage 1.2V). They are not suitable for this product, can damage the monitor and will cause inaccurate readings to be obtained.
pressed and released, the YEAR will stop blinking and the MONTH will begin to blink. Press and release the “ ” button until the desired month is reached. See Figure 5. Repeat this process to set the DAY, HOUR, MINUTES. After you change the batteries, you have to readjust the date and time. Time is maintained using a 24 HOUR clock. AM/PM is not displayed. NOTE: When the number that you wish to set – i.e.
★ Keep up right position on the same height of heart. ★ Do not bend with the cuff or the air tube. Do not inflate before fitting the cuff. ★ When the cuff is dirty, detach it from the equipment, wash the cuff by hand with proper detergent and rinse it in the enough cold water, dry in air. Never iron it. ★ Only use the manufacturer cuff with the main unit to ensure accurate measurement. NOTE: ★ REFER TO THE DIAGRAM PRINTED ON THE CUFF FOR PROPER PLACEMENT.
Figure 7 After you are in a comfortable position, press the “ON/OFF” button. The device will perform a self verification/check. During this verification/check the LCD will display all “8’s”. At the conclusion of the verification/check the LCD will display “00”. See Figure 8. If the device has voice function, it will speak out the displayed blood pressure, heart rate. If an irregular heartbeat is detected, the IRREGULAR HEARTBEAT symbol will appear and blink in the display screen. See Figure9.
★ When the monitor is turned off, press and release the “ or ” button. LCD will display “AVg” in the upper corner of the LCD Display. The result that is first displayed and the result that is first announced – if in “TALKING” mode – is the average of your latest three measurements. ★ To review the results that are in memory – PRESS the “ or ” button to scroll through previous measurements. Each time you press and release the “ or ” button the next oldest result will be displayed.
Figure 10 From the above figure, we can see the classification of blood pressure for adults is as below. The WHO BP Classification Indicating Bar would show out the blood pressure level by the color indicator .
Figure 11 Shut Down After measurement, press button “ON/OFF” to turn off the device. The device will be automatically power off after 1 minute of none use. Voice Function The device with voice function can speak out in the following state: The device will speak out the prompt that “keep silence to take a cuff at the same height with your heart ” when the measure begin.
★ Activate the blood pressure monitor which with UART port and start testing according to the normal blood pressure monitor operation method. ★ After testing the result including systolic pressure, diastolic pressure and pulse will display on LCD, press the SEND button to send these data to the receiving device such as mobile phone. ● The UART port connector & Cable specification: 1. Any type of USB (micro USB, mini USB, standard USB) or serial connector is defined by the customer. 2. Cable : OD 3.
Figure 1 After mobile phone Bluetooth module searched to find the Bluetooth blood pressure monitor, it display “ClinkBlood” . Ⅱ.Press the “ClinkBlood” till “Connected” display on mobile phone screen which means successfully connected with blood pressure monitor. Ⅲ.Slide mobile phone screen to "Slave - > Host" and switch its NOTIFIED VALUES status from "Listen for notifications" to "Stop listening", which means it will be status of "Listening" once open.
Ⅴ. Press command of “Connect”, then press “Start measuring” command,blood pressure monitor start to measure, open“Log” to see the blood pressure monitor dynamic measurement process. It is hexadecimal. The above two models have all function as same normal NIBP, only have different “result output “method.
V. Troubleshooting Abnormality LCD shows Low Battery icon Shows abnormal result Reason Checkout Batteries are low. Change new batteries. Cuff is not tightened properly or its position is incorrect. The arm is moved during measuring. Irregular heartbeat Shows abnormal result After power on ,no display on LCD Cuff inflation rate is too low or does not inflate Cuff deflates too quickly Tighten cuff correctly and refer to “Arm Cuff Connecting”. Stay calm, arm remains steady.
STATEMENTS AND DECLARATIONS: 1. MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in the ACCOMPANYING DOCUMENTS 2. Wireless communications equipment such as wireless home network devices, mobile phones, cordless telephones and their base stations, walkie-talkies can affect this equipment and should be kept at least a distance d = 3,3 m away from the equipment. (Note.
RF emissions CISPR 11 RF emission CISPR 11 Harmonic emissions IEC 61000-3-2 Voltage fluctuations/ emissions IEC 61000-3-3 guidance The [EQUIPMENT or SYSTEM] use RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
for 25 cycles Power frequency (50Hz) magnetic field IEC 61000-4-8 NOTE <5% UT (>95% dip in UT) for 5 sec 3A/m or a battery. 3A/m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment. UT is the a.c. mains voltage prior to application of the test level. Guidance and manufacture’s declaration – electromagnetic immunity The [EQUIPMENT or SYSTEM] is intended for use in the electromagnetic environment specified below.
b mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the [EQUIPMENT or SYSTEM] is used exceeds the applicable RF compliance level above, the [EQUIPMENT or SYSTEM] should be observed to verify normal operation.
Explanation of Symbols: LOT Symbol for batch code Symbol for manufacturer Symbol for ‘CE” Symbol for “electrical and electronic equipment” Symbol for “TYPE BF APPLIED PART” Symbol for “Follow operating instructions” IP21 Symbol for “the IP classification” Symbol for “ RF transmitters” Manufacturer: FUDAKANG INDUSTRIAL CO.,LTD Address :No.8 Yinghe Road, Yuanjiangyuan Management Zone, Changping Town, Dongguan, Guangdong China. Tel: 86-769-81098181 Software Version1.3 Manual Version:V2.