User's Manual

- 7 -
Description Meaning
Measuring temperature over 43.0℃/109.4
Measuring temperature lower than 34.0℃/93.2
Environment temperature high than
working temperature.10.0- 40.0(50.0- 104.0)
9. Specifications
Model No.
FT-F11, FT-F21, FT-F12, FT-F22; FT-F11-BT, FT-F21-BT;
FT-F12-BT, FT-F22-BT; UO8080, UO8080E
Measuring Range
Body 34.0-43.0( 93.2-109.4)
Measuring time
minimum measuring time is 2 seconds; and minimum time between
measurements for the same site is 2 seconds
Accuracy
±0.2/0.4 34.0-34.9 (93.2- 94.82)
±0.2/0.4 35.0-42.0 (95.0-107.6)
±0.2/0.4 42.1-43.0 (107.78-109.4 )
Resolution
0.1/0.2
Operation Condition
10.0-40.0(50.0- 104.0)
Rh95%
Pressure: 86~106 kPa
Storage Condition
-25.0-55.0(-13.0- 131.0)
Rh95%
Pressure: 86~106 kPa
Battery type and battery life
AAA 2pcs Suitable for FT-F11 FT-F21 ,FT-F11-BT FT-F21-BT,
UO8080, UO8080E
1 PC 2032 button battery Suitable for FT-F12FT-F22 ,FT-F12-BT
FT-F22-BT
The typical service life of the new and unused batteries is 300
measurements for the operation time is 60s
Protection against electric
shock
Internally powered ME equipment
Protection against electric
shock
Type BF
Cuff
IP classification
IP22 (
IP22: The first number 2: Protected against access to hazardous parts
with a finger, and the jointed test finger of 12 mm Ф, 80 mm length, shall have
adequate clearance from hazardous parts .And protected against solid foreign
objects of 12,5 mm Ф and greater. The second number: Protected against
vertically falling water drops when enclosure titled up to 15º. Vertically falling
drops shall have no harmful effects when the enclosure is titled at any angle
up to 15’’ on either side of the vertical.