User's Manual

3-3
Basic Operation
3
DR-ID 900PU Operation Manual 897N102280A
3.1.4 Inserting/Removing the Flat Panel Sensor into/from the
Radiographic Examination Stand
[1] In the case of the at panel sensor DR-ID 911SE
The DR-ID 911SE cannot be mounted to any radiographic examination stand other than one
specied by FUJIFILM Corporation.
For details, refer to the operation manual of the radiographic examination stand.
[2]
In the case of the at panel sensors DR-ID 601SE/
DR-ID 602SE/DR-ID 611SE/DR-ID 613SE
Follow the procedure below to insert/remove the at panel sensor into/from the radiographic
examination stand.
For details, see the Operation Manual for the radiographic examination stand
CAUTIONS
For the positioning at the time of inserting/removing the at panel sensor, see the Operation
Manual for the radiographic examination stand.
CAUTIONS
Be careful not to have your ngers caught when inserting/removing the at panel sensor into/
from the radiographic examination stand.
[1] Upright type
When inserting from the right-hand side
CAUTIONS
When inserting the at panel sensor into
the radiographic examination stand, direct
the exposure plane toward the X-ray tube.
1 Pull out the tray.
Tray
2
Insert the at panel sensor into the cassette
receive while the connector directed to the
upper right, and then move it downwards.
Connector
Cassette receive
3
Set the at panel sensor to the upper part of the tray.
4 Push the tray back into place after setting
the at panel sensor.
When inserting the at panel sensor from the left-
hand side, direct the connector to the lower left.
Connector
5 Remove the at panel sensor after use.
Pull out the tray, push the cassette receive
downwards, and then remove the at panel sensor.
Push the tray back into place.