User's Manual

A-2
Appendix A Specifications
Flat Panel Sensors Operation Manual 897N200608
A.1.4 Image Performance
Each at panel sensor complies with IEC 62220-1 (MEDICAL ELECTRICAL EQUIPMENT -
CHARACTERISTICS OF DIGITAL X-RAY IMAGING DEVICES - ) as a general X-ray radiography
equipment.
To ensure optimal image quality, it is recommended that you do not use the at panel sensor near
devices (motor, transformer, switching supply, etc.) that generate electromagnetic noise.
A.1.5 Load Restriction
Entire surface load :
RIC 35G, RIC 43G, RIC 35C, RIC 43C, RIC 24C :
300kg (661.5 lb)
Local load :
RIC 35G, RIC 43G, RIC 35C, RIC 43C, RIC 24C :
120kg (264.6 lb) / ø40mm (1.6in.) (Based on INNOLUX measurement specications)
CAUTIONS
Do not apply an excessive force to the exposure plane.
The sensor inside the at panel sensor may be damaged, and it may not be possible to make
an exposure properly.
A.1.6 Radio Waves
The at panel sensor uses the following types of radio waves.
Flat panel sensor
Wireless specification IEEE802.11n
Transmit frequency 5.2, 5.3, 5.6, 5.8, 2.4 GHz
Modulation OFDM
Frequency tolerance ±20 ppm
Data transfer rate 35 Mbps
Transfer power 17 dBm or less
CAUTIONS
Transmit frequencies available vary, depending on the country.
Radio waves available outdoors vary, depending on the country where the system is used.
When the RIC and any other wireless equipment are operating on the same frequency
channel in a hospital, it may take time to show an image on the image processing unit
monitor.