ON/OFF of Parameter Display
1. Press the Display ON/OFF key. The confirmation display for ON/OFF of RESP display
will appear.
2. Select Display ON or Display OFF .
Pressing the Display ON key will display the
waveform and numeric data.
Pressing the Display OFF key will not display the
waveform and numeric data.
When ECG electrodes are attached to the patient with the respiration display set to OFF, the
respiration waveform and numeric data will be automatically displayed after 30 seconds.
CAUTION
When waveform and numeric data display is set to OFF, the alarm
generation and tabular/graphic trend will be also set to OFF.
NOTE
If RR source is set to other than Impedance , selecting Display OFF
will display CO
2
_RR for the RESP parameter key.
The Display OFF message will be
displayed inside the parameter key.