User Manual

6
21
BP Alarm (BP1, BP2)
1. Press the BP Alarm key to display the alarm setup menu.
Select ON/OFF of BP alarm and set the upper and lower alarm limit for systolic (SYS), diastolic
(DIA), and mean (MEAN) BP.
The alarm value is to be set for each measurement unit. (mmHg / kPa)
The adjustable increment for upper and lower limit changes from 50mmHg / 7kPa.
mmHg : 0
50mmHg / 2mmHg increment
55
300mmHg / 5mmHg increment
kPa : 0
7kPa / 0.2kPa increment
7.5
40.0kPa / 0.5kPa increment
Key Item Description
Individual Alarm
Selecting ON will generate BP alarm.
Selecting OFF will not generate BP alarm.
Select from SYS (systolic BP), DIA (diastolic BP),
MEAN (mean BP).
Lower Alarm Limit
Sets the lower alarm limit (0
295mmHg/0
39.5kPa).
Setting a value equal to or below 0mmHg/0kPa will
turn OFF the alarm.
Upper Alarm Limit
Set the upper limit (2
300 mmHg/0.2
40.0kPa).
Setting a value equal to or above 300 mmHg/40.0kPa
will turn OFF the alarm.
Automatic Setup
Automatically sets the upper limit to
40mmHg/
5.5kPa, and the lower limit to
20mmHg/
2.5kPa to
the current value.
Zero Balance of Pressure Lines (BP1, BP2)
1. Open the three-way cock of the pressure transducer to air.
2. Press the BP zero key.
6
BP