User Manual

6
23
CAUTION
If the mean BP display is set to OFF, the mean BP alarm will not be
generated. Also, the mean BP will not be displayed on the tabular trend.
Be cautious when setting the mean BP display OFF.
ECG Source (BP1)
The HR/PR source to display on the home display can be selected.
The alarm will be generated based on this selection.
The graphic trend and tabular trend will be also stored based on this selection.
BP2 can not be set as ECG source.
1. Press the Configuration
key to display the setup menu to set the ECG source.
2. Select a parameter.
Selecting ECG will measure the HR from ECG.
HR will be displayed inside the parameter key.
Selecting SpO
2
will measure the PR from SpO
2
.
PR_SpO
2
will be displayed inside the parameter key.
Selecting BP1 will measure the PR from BP1.
PR_BP will be displayed inside the parameter key.
Selecting Auto
will automatically set the measurable HR source in the priority of ECG
SpO
2
BP1.
6
BP