User Manual

6
25
SpO
2
This menu allows setup of the measurement condition for the SpO
2
.
Size : Sets the waveform size for SpO
2
waveform display.
SpO
2
Alarm : Sets ON/OFF of alarm, upper and lower alarm limit, and SEC alarm.
Configuration : Sets the SpO
2
monitoring configuration.
CAUTION
Take the following precautions when monitoring over long periods of time.
?
To avoid skin rash or low-temperature burn, it is recommended to
change the measurement position several times a day.
Be especially careful when continuously using on neonates, infants, or
patients with peripheral circulatory disturbance.
?
Direct sunlight to the sensor area can cause a measurement error.
Place a black or dark cloth over the sensor in these environments.
When not measuring, unplug the relay cable and sensor from the SpO
2
connector. Otherwise, the outside light may affect to falsely display
measurements.
CAUTION
?
The DS-100A is intended for use on finger of adults weighing over 40
kg (approximate). Do not use them on children or neonates. Also do
not apply them on the thumb or foot.
?
The light-emitting part of the sensor should be over the root of the
fingernail. Do not insert the finger too far into the sensor as it may hurt
the patient.
?
The DS-100A is not designed for long term use. Remove the sensor
every 4 hours. If any inhibition is detected in tissue blood flow, replace
it or move the sensor to another finger.
?
Measuring on a limb with NIBP cuff, arterial catheter, or intracatheter
may result in incorrect measurement.
?
Do not secure the adhesive tape too tight as it may obstruct the blood
flow.
6
SpO
2