ECG Source
The HR/PR source to display on the home display can be selected.
The alarm will be generated based on this selection.
The graphic trend and tabular trend will be also stored based on this selection.
1. Press the Configuration
key to display the setup menu to set the ECG source.
2. Select a parameter.
Selecting ECG will measure the HR from ECG.
HR will be displayed inside the parameter key.
Selecting SpO
2
will measure the PR from SpO
2
.
PR_SpO
2
will be displayed inside the parameter key.
Selecting BP1 will measure the PR from BP1.
PR_BP will be displayed inside the parameter key.
Selecting Auto
will automatically set the measurable ECG source in the priority of ECG
SpO
2
BP1.
SpO
2
Alarm during NIBP Measurement (Ignore NIBP)
This setup is to be made when the SpO
2
sensor and NIBP cuff is placed on the same limb for
measurement.
During the NIBP measurement, the cuff inflation restricts the blood flow which disables the correct
detection of the SpO
2
value and PR, and may generate an improper alarm.
Selecting OFF
will not generate the alarm until the NIBP measurement is complete. Similarly,
when the HR source is set as SpO
2
,
the PR alarm will not be generated during NIBP
measurement.
1. Press the Configuration key to display the setup menu for setting Ignore NIBP.
2. Select ON or OFF .
ON
will generate the alarm during NIBP measurement.
OFF
will not generate the SpO
2
/PR alarm during NIBP measurement.