Non-Invasive Blood Pressure
This menu allows the setup of NIBP monitoring condition.
Auto Mode : Sets the automatic interval measurement and starts the 1-minute interval
measurement and Quick SYS measurement.
NIBP Alarm : Sets the ON/OFF of NIBP alarm and upper / lower limit of systolic, diastolic, and
mean BP.
Configuration : Sets the NIBP monitoring configuration.
CAUTION
For the following situation, measurements will be terminated.
?
When the measurement time has exceeded 120 seconds for adult, 90
seconds for child, 60 seconds for neonate.
?
When the inflation value has exceeded 300mmHg for adult, 200mmHg
for child, 150mmHg for neonate.
CAUTION
If used with the incorrect patient type, it will not only cause erroneous
measurement, but the inflating level for the adult may be applied to child or
neonate causing a dangerous situation to the patient.
NIBP Automatic Measurement
Non-invasive blood pressure can be measured automatically at selected time intervals. If Quick
SYS measurement is performed during the NIBP automatic measurement, the automatic
measurement will automatically resume when Quick SYS measurement completes.
When NIBP automatic measurement is set, the set interval time
will be displayed inside the parameter key.
1. Press the Auto Mode key to display the measurement interval setup menu for the
automatic measurement.
Currently selected time interval