User Manual

6
30
2. Select an interval time.
Select from 2 min
/ 2.5 min / 3 min / 5 min / 10 min / 15 min / 20 min /
30 min
/ 60 min / 120 min .
Select OFF
if not performing the interval measurement.
The measurement time will be the integral multiple of the selected interval time beginning with 0
minute.
Ex.
If the present time is 13:14, the measurement time will be as follows for each interval time.
2 min. : 13:16, 13:18, 13:20,
...
2.5 min. : 13:15, 13:17:30, 13:20,
...
3 min. : 13:15, 13:18, 13:21,
...
5 min. : 13:15, 13:20, 13:25,
...
NIBP 1-Minute Interval Measurement
The 1-minute interval measurement will automatically stop after 10 minutes or 20 minutes and
returns to the previous interval mode setup.
1. Press the Auto Mode
key to display the measurement interval setup menu to start the
1-minute interval measurement.
2. Press the 1min Start key to start the 1-minute interval measurement.
Pressing the NIBP START/STOP
key will not stop the 1-minute interval measurement.
To cancel the measurement, press the 1min Start
key again.
The measurement duration of 1-minute interval measurement can be selected on the 1min Auto
of the NIBP configuration menu.
10min.
will automatically stop the 1-minute interval measurement after 10 minutes and returns
to the previous interval mode setup.
20min.
will automatically stop the 1-minute interval measurement after 20 minutes and returns
to the previous interval mode setup.
CAUTION
?
The 1-minute interval measurement will always start from 00 second.
Pressing the 1min Start
key will start the measurement from the
next 00 second.
?
The 1-minute interval measurement will automatically stop after 10
minutes or 20 minutes and returns to the previous interval mode setup.
The selection of 10min
/ 20min can be made on the NIBP
configuration menu. (Refer to 1-Minute Measurement Duration of
this section
Measurement Duration