User Manual

6
32
Key Item Description
Individual
Alarm
Selecting ON
will generate the NIBP alarm.
Selecting OFF
will not generate the NIBP alarm.
Select from SYS (systolic BP), DIA (diastolic BP), or
MEAN (mean BP)
Lower Alarm
Limit
Sets the lower alarm limit (10
295mmHg/1.5
39.5kPa).
Setting a value 10mmHg/1.5kPa or below will turn OFF
the alarm.
Upper Alarm
Limit
Sets the upper limit (15
300mmHg /2.0
40.0kPa).
Setting a value 300bpm/40.0kPa or above will turn OFF
the alarm.
Automatic
Setup
Automatically sets the upper limit to
40mmHg/
5.5kPa to the current value, and the lower limit to
20mmHg/
2.5kPa to the current value.
Quick SYS Measurement Duration
The duration of Quick SYS can be selected from 3 min., 5 min., or 10 min.
The long duration of continuous measurement may congest the blood stream of the measured
location. Set the duration according to the patient condition.
1. Press the Configuration
key to display the NIBP configuration menu to set the Quick
SYS.
2. Select the measurement duration.
Select an appropriate time from 3min
, 5min , 10min .
Quick SYS will automatically cease after the selected duration.