User Manual

6
4
ECG
The measurement condition for ECG can be set on this menu.
Lead, Size : Sets the waveform size and lead for ECG display and recording.
HR Alarm : Sets ON/OFF of HR alarm, and sets upper and lower alarm limit.
Arrhy. Alarm : Sets ON/OFF and detection threshold for each arrhythmia alarm.
Configuration : Sets the condition for measuring ECG and HR.
Arrhy. Learn : The monitor learns the normal QRS at ECG electrode replacement or at
misdetection of the arrhythmia analysis.
ECG Waveform Size and Lead
1. Press the Lead, Size key to display the size / lead setup menu.
2. Select ECG1 or ECG2.
If 4-electrode or 5-electrode ECG relay cable is used,
2 channels of ECG can be measured.
Select ECG1
or ECG2 key to set the
waveform size, lead, baseline position. The key
LED for the selected channel will light.
When 3-electrode is used, these keys will not be
displayed.
3. Select the waveform size.
Select the waveform size for displaying and
recording.
Size
×
1/4
×
1/2
×
1
×
2
×
4
Voltage (10mm) 4mV 2mV 1mV 500uV 250uV