User Manual

6
40
EtCO
2
Average Duration
The duration to average the EtCO
2
value can be selected from 10 sec., 20 sec., 30 sec., or OFF.
1. Press the Configuration key to display configuration menu for EtCO
2
average duration
selection.
2. Select the average duration.
Select the duration to average the EtCO
2
value from 10sec
, 20sec , 30sec .
If OFF
is selected, EtCO
2
value for each respiration will be displayed.
As the EtCO
2
value display is updated each second, EtCO
2
value for each respiration can not be
displayed if respiration rate is above 60 Bpm.
Measurement Unit
The measurement unit can be selected from mmHg, kPa, or %.
Measurement Unit
1. Press the Configuration key to display the configuration menu for measurement unit
selection.
2. Select the measurement unit from mmHg
, kPa , % .
The graphic trend and tabular trend will be displayed with the selected measurement unit.
If the measurement unit is changed frequently, the continuity of the graphic trend and tabular trend
may be lost.
When the measurement unit is changed, make sure to set the alarm condition for that unit. The
alarm setup is necessary for each measurement unit.