User's Manual

2
7
Preparation for Monitoring To Turn On the Power
Connecting the Power Cable
Connect the accompanying AC power cable (CS-34) to the monitor and to the 3-way grounded
outlet.
When the AC power is supplied to the monitor, the main power lamp will light in green.
When the battery pack (optional) is installed, the charge lamp will light and charging will start.
WARNING
?
Use only the accompanying 3-way AC power cable. Use of other
cables may result in electric shock to the patient and the operator.
?
The power cable must be connected to AC 115V hospital grade outlet.
?
When using multiple ME equipment simultaneously, perform
equipotential grounding to prevent potential difference between the
equipment. Even a small potential difference may result in electric
shock to the patient and the operator.
Power Supply
Connector
Charge Lamp
This lamp will light in orange during charging.
It will turn to green when the charging is complete.
Main Power Lamp
This lamp will light in green when the AC power is supplied.
The lamp will be extinguished during battery operation.