User's Manual

2
9
To Use with the Battery Pack
Unplug the power cable from the power supply
connector located at the rear side of the monitor.
If the power cable is unplugged with the power turned
ON, the monitor will also automatically change to
battery operation mode.
Turn ON the power switch located at the right side of
the monitor.
Indication of Operation Time
Battery Mark
Charge
Condition
Standard Mode Power Saving Mode
Full
3 hours
2 hours 20 min. 3 hours 30 min.
2 hours 40 min.
Medium
2 hours
10 min. 2 hours 40 min.
10 min.
Empty
10 min, 10 min.
Reference
Refer to 8. System Configuration Monitor Setup for power saving mode.
Indication of Charging Time (Empty to Full)
When the power cable is connected and the
power is turned OFF
2.5 hours
When the power cable is connected and the
power is turned ON
13 hours
CAUTION
The above operation time indicates the time for the new battery pack
performing ECG measurement, NIBP periodic measurement (5-minute
interval). Note that the battery pack degrades with a continuous use and
shortens the usable time.
Power Supply
Connector
Power Supply Switch
During battery operation, a battery mark will be displayed on the
upper part of the display.
2
Preparation for Monitoring