(3) Inject the heparin 1000 units to saline bag, mix
thoroughly and puncture the infusion line through
the same hole. Set the saline bag to pressure
bag, and hang from the infusion device. Fill
saline to about 1/3 of the drip.
(4) After loosening the zero-port plug, push the flash
button to perform priming to remove air bubbles.
(5) Verify that all air bubbles are removed, and tighten
the zero-port plug.
Turn on the zero-port plug side of the open-air
three-way cock.
(6) Inflate the pressure bag to 300mmHg.
(7) Wait about 5 minutes after setting the blood
pressure device.
Next procedure is to perform the zero balance.