Description of Alarm Message and Alarm Sound
This section explains about the message displayed on the home display.
There are vital alarm message and equipment status alarm message which will be displayed on
the top of the home display.
The alarms are classified in level 1, level 2, level 3, level 4, and the alarm message will be
displayed according to the priority of level 1
level 2
level 3
level 4.
The color of the displayed messages are red for level 1, yellow for level 2, blue for level 3, and
white for level 4.
Alarm Level Description Tone Displayed Color
Level 1 Life Threatening Alarm Continuous beep tone Red
Level 2 Cautionary Alarm Beep tone every 5 seconds Yellow
Level 3 Treatment Needed Alarm Single beep tone Blue
Level 4 Notification Alarm Display Only White
CAUTION
?
Alarm messages will be displayed according to the priority. (Level 1
Level 2
Level 3
Level 4)
?
For the same alarm level, the alarm message for the newer alarm will
be displayed.
Vital Alarm Message
The vital alarm message is generated when a measurement exceeds the alarm limit, or when
arrhythmia is detected.
There are 2 types of alarm messages, numeric alarm message and arrhythmia alarm message.
If the 2 types of alarm generate at the same time, the numeric alarm message and arrhythmia
alarm message will be alternately displayed for 2 seconds each. The message will be displayed
according to the priority of the alarm level. If the alarms of the same level generate, the message
for the newer alarm will be displayed.
CAUTION
The alarm message for the arrhythmia alarm will continue to be displayed
for 30 seconds after the alarm is resolved.
Numeric Alarm Message
Arrhythmia Alarm Message