User Manual

4
22
5. Start the periodic recording.
ON will activate the periodic recording with the
selected interval. If periodic recording is not
necessary, select OFF
.
Alarm Recording
The recording will automatically start at occurrence of numeric alarm or arrhythmia alarm.
CAUTION
If the alarm with the higher priority occurs during alarm recording, the
recording in process will be ceased and starts the alarm recording with the
higher priority.
1. Press the Menu
System Config.
Record
Alarm Record
keys.
The alarm recording setup menu will be displayed.
2. Select the waveform for recording.
Up to 3 waveforms can be selected. The waveforms
will be automatically located at recording.
Alarm
will record the waveform which generated
the alarm.
3. Select the recording factor.
Select the recording factor for alarm recording.
HR will start the alarm recording when the HR or
PR alarm is generated.
Other
will start the alarm recording when the
numeric alarm other than HR and PR alarm is
generated.
Arrhy.
will start the alarm recording when the
arrhythmia alarm is generated.
4. Select the recording duration.
Select the recording duration from 12sec ,
24sec . The recording will automatically stop after
the selected time.