The delay time differs depending on the recording time.
Delay Time
Neonate
Recording
Time
Adult Child
Numeric Alarm Arrhythmia Alarm
12 sec. 12 sec. 12 sec. 8 sec. 12 sec.
24 sec. 16 sec. 16 sec. 16 sec. 16 sec.
NOTE
5. Select the arrhythmia type.
If arrhythmia is selected for the recording factor, select the arrhythmia type.
Pressing the Arrhythmia Record key will display
the arrhythmia selection window for alarm recording.
Arrhythmia Recording Setup Menu
Selected as alarm recording factor.
Not selected as alarm recording factor.
6. Start the alarm recording.
ON will automatically start the recording at alarm
occurrence.
If alarm recording is not required, select OFF .
Center will start the alarm recording on the central
monitor with the smallest monitor ID.
NOTE
The numeric data to be recorded will be the data at alarm occurrence time.