User Manual

4
25
Graphic Recording (Graphic/Tabular Trend, etc.)
Graphic trend recording, tabular trend recording, NIBP list recording, and recall recording can be
performed.
Reference
Refer to 7. Function Graphic Trend, Tabular Trend, Recall Data, NIBP List for recording
procedure.
Graphic Trend
One parameter from the following can be selected for graphic recording.
Parameter Description
HR HR, PR (SpO
2
, BP)
ST ST1, ST2
VPC VPC beats
BP BP (SYS / Mean / DIA)
NIBP NIBP (SYS / Mean / DIA)
SpO
2
SpO
2
value
PR SpO
2
pulse rate
TEMP Temperature
RR Respiration Rate (Impedance, CO
2
)
APNEA Apnea Time (Impedance, CO
2
)
CO
2
EtCO
2
/ InspCO
2
EVENT1 ASYSTOLE, VF, VT, SLOW_VT, RUN, BIGEMINY
EVENT2
TRIGEMINY, PAUSE, COUPLET, TACHY, BRADY,
FREQUENT
Tabular Trend
12 parameters from the following can be displayed / recorded as tabular trend.
Parameter Description
HR HR, PR (SpO
2
, BP)
PR_SpO
2
PR (SpO
2
)
VPC VPC beat
ST1 ST level of ECG1
ST2 ST level of ECG2
RR RR (Impedance, CO
2
)
APNEA Apnea Time (Impedance, CO
2
)
SpO
2
SpO
2
value
BP1 BP(SYS / Mean / DIA)
BP2 BP(SYS / Mean / DIA)
NIBP NIBP(SYS / Mean / DIA)
TEMP Temperature
EtCO
2
EtCO
2
value
InspCO
2
InspCO
2
value
NIBP List
The numeric data for the following parameters can be recorded.
Parameter Description
HR HR, PR (SpO
2
, BP)
PR PR (SpO
2
)
SpO
2
SpO
2
value
NIBP NIBP (SYS / Mean / DIA)
Recall
The waveform and numeric data at alarm occurrence will be recorded with the following setup.
Parameter Description
Waveform (Max. 2) ECG1, ECG2, BP, SpO
2
, CO
2
, RESP
Numeric Data
HR, ST, RR, TEMP, NIBP, BP, ST1, ST2, SpO
2
,
PR_SpO
2
, APNEA, CO
2
Recording Duration 12 sec.
4
Recording Setup