User Manual

4
26
Recorder Operation (QRS Symbol, etc.)
This menu allows the setup of paper feed operation and ON/OFF of QRS classification symbol
recording
1. Press the Menu
System Config.
Record
Setup
keys.
The recorder operation setup menu will be displayed.
2. Select ON/OFF of paper feed to top.
ON will feed the paper to start recording from the
next perforation.
OFF will not feed the paper and starts recording
from the position where the previous recording
ended.
3. Select ON/OFF of paper feed to end.
ON will feed the paper after recording to next
perforation so that the paper can be easily cut off.
OFF
will not feed the paper after recording.
4. Select ON/OFF of QRS classification symbol recording.
ON will record the QRS classification symbol on to
the ECG waveform.
Symbol Description
N (Normal) Normal QRS beat
V (VPC) Ventricular Extrasystole
S (SVPC) Supraventricular Extrasystole
P (Pacing Beat) Pacing beat
F (Fusion Beat) Fusion beat of pacing and spontaneous beat.
? (Undetermined beat) Learning arrhythmia, or beat not matching the pattern
NOTE
?
The QRS classification symbol can not be recorded for the manual recording
without delay time, and for the periodic recording.
To record the QRS classification symbol, set the delay time to 8 seconds or 16
seconds for manual recording.
?
For the freeze recording, the QRS classification symbol may not be recorded if
the recording is started immediately after the waveform display is frozen.
In such case, start the recording after about 10 seconds from the time the
waveform is frozen.