User Manual

4
31
5. Enter the group ID.
Press the Group key, and enter the group ID.
Use the numeric keypad to enter the group ID in the
range from 00 ~ 63.
6. Save the telemetry channel ID and group ID.
Press the Save key to store the channel ID and
group ID. Verify that the Complete message is
displayed.
If an error is found on the password, channel ID, or group ID, the following message will be
displayed.
Invalid password : The entered password is incorrect. Enter the password
again and press the Save key.
Invalid data : The entered channel ID or group ID is outside the
programmable range. Enter the ID within the range and
press the Save
key.
Telemetry unit error : Abnormality is found on the telemetry transmitter. Contact
our service representative.
Install Program Card : Install the program card to PC card slot.
7. Verify the stored telemetry channel ID and group ID.
NOTE
When the monitor indicates that the measurement data is out of range
(“
× × ×
display), the minimum or maximum value of the range will be displayed
at the telemetry center.
Out of range
Telemetry Center
HR 301bpm or above 300bpm
RR 151bpm or above 150bpm
BP -51mmHg or below -50mmHg
301mmhg or above 300mmHg
TEMP -0.1
?
C or below 0
?
C
50.1
?
C or above 46.1
?
C
*For temperature, the measurement range is up to 50.1
?
C. But at the
telemetry center, 46.1
?
C will be the maximum value displayed.
CAUTION
It is necessary to insert the program card to the PC card slot to set the
channel ID and program ID.
4
Telemetry Setup