Transmitting Waveform Setup
This function allows to select the transmitting waveform.
1. Press the Menu
System Config.
Telemetry Wave Setup
keys.
2. Select the waveform to transmit.
Select the waveform to transmit from ECG1 ,
ECG1+2 .
Selecting ECG1 will transmit ECG1, CO
2
, RESP,
BP1, BP2, SpO
2
.
RESP waveform will be transmitted
when Impedance is selected for RR source.
Selecting ECG1+2
will transmit ECG1, ECG2,
CO
2
/RESP, BP1, SpO
2
.
CO
2
and RESP waveform to be transmitted will be in
accordance with RR source setup.
NOTE
The waveform not displayed on the home display will not be transmitted
even if selected as the transmitting waveform.
CAUTION
?
If the temperature measurement unit is
?
F, or BP measurement unit is
kPa, the corresponded waveform and numeric data will not be
transmitted.
For telemetry transmission, use
?
C for temperature unit and mmHg for
BP unit.
?
For telemetry transmission, BP waveform with a scale above the
programmed scale can not be properly transmitted. Set the BP
waveform below the programmed scale.