User Manual

4
40
To Turn ON the System Alarm
Turning the system alarm ON will activate the alarm setups for each parameter.
1. Press the Menu
Alarm
keys.
2. Set the alarm ON
.
Press the Suspend
key when the key LED is lighted. The key LED will extinguish.
The alarm is ON when the key LED is extinguished.
To Suspend the System Alarm
The alarm can be temporarily suspended. During the alarm suspension, Alarm Suspended
**
sec. message will be displayed.
***
sec. indicates the remaining suspended time. The
alarm will turn ON when the suspended time completes.
1. Press the Menu
Alarm
keys.
2. Suspend the alarm.
Press the Suspend key when the key LED is extinguished. The key LED will light.
Alarm is suspended when the key LED is lighted.
To Silence the System Alarm
The alarm sound can be silenced for fixed amount of time. This setting will not affect the alarm
message. If the alarm cause still remains at completion of silence time, the alarm sound will
generate again. Also, if another alarm with the same or higher priority occurs during the alarm
silence time, the alarm sound for the new alarm will generate.
Precautions about Silencing the Alarm
?
Alarm silence function is effective for each parameter. If an alarm condition for the selected
parameter is resolved for a moment but is generated again during the alarm silence time, the
alarm will remain silenced. The recall and alarm recording will not function at this time.
?
If another alarm with the lower priority occurs during the alarm silence time, alarm sound will not
generate. The recall and alarm recording will function.
?
If the Alarm Silence
key is pressed for the alarm of another parameter which occurred during
the alarm silence time, the alarm silence time for the first alarm will not be extended.
?
The alarm silence condition for all parameters will be ceased in the event of any of the following.
When the main power is turned ON.
When the system alarm status (ON / suspend) is changed.
When monitoring is suspended on the patient admit / discharge menu.
When the alarm mode is changed on the patient admit / discharge menu.
When the patient has discharged.
?
The alarm silence condition for each parameter will be ceased in the event of any of the
following.
When the alarm silence time for the parameter is completed.
When automatic alarm is selected for the parameter.
When the alarm is turned OFF for the parameter.