Alarm Setup for Each Parameter
The alarm for each parameter can be turned ON or OFF, and upper and lower alarm limit can be
set.
1. Press the Menu
Alarm
keys.
The alarm setup menu will be displayed.
2. Select the parameter to display the alarm setup menu.
Display Description
Displays the upper and lower alarm limit and the current
measurement (
).
The limits can be adjusted by directly pressing the bar
display or using the arrow keys.
Displays lower limit
current value
upper limit.
Key Item Description
Individual Alarm
Selecting ON
will generate the alarm.
Selecting OFF will not generate the alarm.
Lower Alarm Limit
Sets the lower alarm limit.
The lower limit will be turned OFF when a value below
the range is selected.
Upper Alarm Limit
Sets the upper alarm limit.
The upper limit will be turned OFF when a value
above the range is selected.
Automatic Setup
Automatically sets the limits corresponding to the
current value.
If the limit is turned OFF, it will be remained OFF.
The system alarm and parameter alarm will be in a
ON condition.