42
3. Select ON/OFF and set upper and lower alarm limit for the parameter alarm.
Numeric Data Key Item Description
HR / PR / BPR ON, OFF 20
300bpm
ST
ON, OFF
ST1
±
2.0mV
ST2
±
2.0mV
BP1 ON, OFF 0
300mmHg
BP2
ON, OFF 0
300mmHg
SpO
2
ON, OFF 50
100%
RR
ON, OFF 5
150bpm (Adult)
2
150bpm (Child, Neonate)
APNEA (Upper Limit)
ON, OFF 5
20 sec.
TEMP ON, OFF 30
50
NIBP ON, OFF 10
300mmHg
EtCO
2
ON, OFF 1
100mmHg
     
0.1
13.3kPa
     
0.1
13.3
InspCO
2
(Upper Limit)
ON, OFF 1
4mmHg
     
0.1
0.4kPa
     
0.1
0.4%