To Set ON/OFF of Arrhythmia Alarm
Selecting ON will generate the arrhythmia alarm.
Selecting OFF will not generate the arrhythmia alarm.
However, Asystole, VF, and VT alarm can not be turned OFF at
the initial factory setting. Select ON/OFF for Asystole, VF,
VT on the Hospital Setup menu to turn OFF the alarm.
Reference
Refer to 8. System Configuration Hospital Setup
Page Item Selection
Asystole ON, (ON, OFF)
VF ON, (ON, OFF)
VT ON, (ON, OFF)
Page 1/3
Slow_VT ON, OFF
Run ON, OFF
Bigeminy ON, OFF
Trigeminy ON, OFF
Page 2/3
Pause ON, OFF
Couplet ON, OFF
Tachy ON, OFF
Brady ON, OFF
Page 3/3
Frequent ON, OFF
NOTE
The Arrhythmia alarm OFF message will be displayed when the ASYSTOLE,
VF, VT, SLOW_VT, and HR alarm is OFF.
To Set the Arrhythmia Detection Level
Select the level to detect each arrhythmia.
1. Pressing the detection level key (ex. 5 for Asystole) for each arrhythmia will display the
window to adjust the detection level.
2. Set the detection level.
Use the arrow keys to set the detection threshold.
Item Range
ASYSTOLE 3
10 sec.
RUN 2
8 beats
PAUSE
1.5
5 sec.
FREQUENT 1
50 beats/min.
3. Close the window to adjust the detection level.
Press the Close key.