User Manual

5
4
Patient ID
Up to 10 characters of alphabets, numbers, or symbols can be used for the patient ID.
Enter the ID according to the monitoring purpose. The entered ID will be output on the recording
paper.
1. Press the ID key.
Enter the ID using the alphanumeric keypad.
Patient Type
The selection of patient type affects the accuracy of NIBP measurement, HR measurement, and
RR measurement. Also the delay time to generate the measurement data alarm will change
according to the patient type.
Adult Child Neonate
NIBP measurement range 10
280mmHg 10
180mmHg 10
120mmHg
HR 0bpm, 12
300bpm 0bpm, 30
300bpm
Filter Mode (Monitor)
0.5
40Hz 1.6
40Hz
Filter Mode (ESIS) 1.6
15Hz 1.6
15Hz
Filter Mode (ST Display) 0.05
40Hz none
Impedance Respiration 1.5Hz 2.5Hz
Alarm delay time 5 sec. 0 sec.
The alarm delay time is the function to prevent frequent generation of the measurement data
alarm by holding the alarm generation for the duration of each delay time.
The alarm delay time applies to the measurement data alarm for the following parameters; HR /
PR, BP, RR, SpO
2
, TEMP, EtCO
2
/ InspCO
2
, TACHY, BRADY.
WARNING
The monitor determines the detection algorithms for QRS and NIBP
according to the selected patient type. Make sure the proper selection
is made.
Patient ID