User Manual

5
9
Monitoring Mode Selection Alarm and Display Mode
This menu allows to select the alarm mode and display mode.
The alarm setting and display configuration can be each selected from 5 modes depending on the
monitoring purpose. Select the appropriate mode when admitting a patient.
CAUTION
The setup for the alarm mode and display mode remains stored even when
the power is turned off or when discharging procedure is performed.
Before monitoring, make sure the current monitoring mode is suitable for
the patient’s condition.
Mode Selection
1. Press the Mode Select key.
The mode selection menu for alarm mode and display mode will be displayed.
2. Select an alarm mode from the selection.
Press one key from the Alm Mode 1 ~
Alm Mode 5
keys to select an alarm setup mode
which meets the monitoring purpose.
The setup for the alarm mode remains stored even
when the power is turned off or when discharging
procedure is performed. The previously selected
alarm mode will be effective if the selection is not
made.
5
Monitoring Mode Selection